YSI 5200 Recirculating System Monitor Specifications


Specs Dissolved Oxygen pH ORP Conductivity Salinity Temp.
Sensor Type Membrane covered steady state polarographic Gel reference system Platinum button with Ag/Agcl as reference 4 electrode, cell constant of 5.0 ± 5% Calculated from temperature and conductivity 10,000 ohm @ 25°C thermistor
Range 0 to 500% air saturation / 0 to 60 mg/l 0.0 to 14.0 pH -1000 mV to +1000 mV 0 to 500 µS,
0 to 5000 µS,
0 to 50 mS,
0 to 200 mS
0 to 80 ppt 0 to 45°C,
32 to 99°F
Software selectable
System Accuracy ± 2% or ± 0.2 mg/l whichever greater (0-200%), ± 6% of reading (200-500%) 0.1 pH ± 10 mV ± 0.5% of range ± 2% or ± 0.1 ppt ± 0.2°C

Power requirements (DC): 12 VDC/8.5 to 16.5 VDC/500 - 800mA

Output relays: 4 ea 5 amp CE and UL rated 240 VAC 50/60 HZ SPDT relays